The Venturi® Ventilator
Technical Specifications

Control Settings

  • Tidal Volume (compliance compensated) 50 to 2,000 mL
  • Breathing Rate 2 to 150 bpm
  • Peak Inspiratory Flow 10 to 120 L/min
  • Oxygen Percentage 21 to 100%
  • Peak Inspiratory Pressure 0 to 120 cmH2O
  • Exhalation Assist 0 to 30 cmH2O/L/s
  • PEEP 0 to 20 cmH2O
  • Inspiratory Time 0.1 to 4s
  • Inspiratory Pause Time 0 to 1s
  • Sensitivity (Patient Effort Trigger) 0 to 250 mL
  • Flow Drop-Off Percentage (Percent of Peak) 5% to 80%
  • Airway Temperature 15 to 40deg. C
  • Humidifier Temperature 35 to 75deg. C
  • Waveform Shape (clinician modifiable) Custom, square, decelerating, modified sine

Alarms and Indicators

  • High/Low Exhaled Tidal Volume
  • High/Low Respiratory Rate
  • Low Oxygen Flow
  • Low Air Flow
  • Low Oxygen Pressure
  • No High Pressure Air
  • High/Low Airway Pressure
  • High/Low Inspiratory Time
  • High/Low Inspiratory:Expiratory Ratio
  • High/Low Minute Volume
  • Airway Leak
  • Patient Disconnect
  • Apnea Alarm/Backup Ventilation
  • Battery in Operation
  • Check Breathing Bag
  • High/Low Oxygen

Pneumatic Requirements

  • Source Pressure
  • Oxygen (Medical Grade) 45-75 psig
  • Air (Medical Grade) 45-75 psig
  • Source Flow
  • Oxygen (Medical Grade) 100 L /min
  • Air (Medical Grade) 100 L /min

Monitored and Displayed Parameters

  • Exhaled Tidal Volume (Compliance Comp) 50 to 2,500 mL
  • Measured Breathing Rate 2 to 150 bpm
  • Peak Inspiratory Flow 10 to 200 L/min
  • Oxygen Percentage 18 to 100%
  • Peak Inspiratory Pressure 0 to 120 cmH2O
  • PEEP 0 to 20 cmH2O
  • Mean Airway Pressure 0 to 120 cmH2O
  • Inspiratory Time 0.2 to 4s
  • Inspiratory:Expiratory Ratio 0.01 to 99.0
  • Minute Ventilation-Controlled 0 to 99 L/min
  • Minute Ventilation-Spontaneous 0 to 99 L/min
  • Airway Temperature 15 to 40 deg. C
  • Lung Compliance 10 to 150 mL/cmH2O
  • Airway Resistance 1 to 60 cmH2O/L/s
  • Leak 0 to 20 L/min
  • Airway Flow Waveform -120 to +120 L/min
  • Airway Pressure Waveform -20 to +60 cmH2O
  • Flow-Volume Graph
  • Pressure-Volume Graph
  • Minute Volume Wheel

Modes

  • The Venturi® Ventilator's flexible controls provide for a full range of mandatory and spontaneous breaths.
  •  Assist Control, Pressure Control, Variable Pressure Control
  • Pressure Support/SIMV, CPAP/Assist Control,
  • Pressure Support/Variable Pressure Control,
  • Variable Pressure Support/Variable Pressure Control,
  • CPAP/SIMV, Variable Pressure Support/SIMV,
  • Apnea Ventilation

Electrical Requirements

  • Voltage 115 VAC, 6 A or 230 VAC, 3 A
  • Leakage Current <300 mA @ 115 VAC
  • Power Cord 125 VAC, hospital grade, UL and CSA approved, 10 ft length.
  • Internal Batteries Lead Acid 24 V, 18 A-h

Dimensions

  • Height 52 in. (132 cm)
  • Width 16 in. (41 cm)
  • Depth 24 in. (61 cm)
  • Weight 170 lbs. (77 kg)