Control Settings
- Tidal Volume (compliance
compensated) 50 to 2,000 mL
- Breathing Rate 2 to 150 bpm
- Peak Inspiratory Flow 10 to 120
L/min
- Oxygen Percentage 21 to 100%
- Peak Inspiratory Pressure 0 to 120 cmH2O
- Exhalation Assist 0 to 30 cmH2O/L/s
- PEEP 0 to 20 cmH2O
- Inspiratory Time 0.1 to 4s
- Inspiratory Pause Time 0 to 1s
- Sensitivity (Patient Effort
Trigger) 0 to 250 mL
- Flow Drop-Off Percentage (Percent
of Peak) 5% to 80%
- Airway Temperature 15 to 40deg. C
- Humidifier Temperature 35 to 75deg. C
- Waveform Shape (clinician
modifiable) Custom, square, decelerating,
modified sine
Alarms and Indicators
- High/Low Exhaled Tidal Volume
- High/Low Respiratory Rate
- Low Oxygen Flow
- Low Air Flow
- Low Oxygen Pressure
- No High Pressure Air
- High/Low Airway Pressure
- High/Low Inspiratory Time
- High/Low Inspiratory:Expiratory Ratio
- High/Low Minute Volume
- Airway Leak
- Patient Disconnect
- Apnea Alarm/Backup Ventilation
- Battery in Operation
- Check Breathing Bag
- High/Low Oxygen
Pneumatic Requirements
- Source Pressure
- Oxygen (Medical Grade) 45-75 psig
- Air (Medical Grade) 45-75 psig
- Source Flow
- Oxygen (Medical Grade) 100 L /min
- Air (Medical Grade) 100 L /min
|
Monitored and Displayed Parameters
- Exhaled Tidal Volume (Compliance
Comp) 50 to 2,500 mL
- Measured Breathing Rate 2 to 150
bpm
- Peak Inspiratory Flow 10 to 200 L/min
- Oxygen Percentage 18 to 100%
- Peak Inspiratory Pressure 0 to 120
cmH2O
- PEEP 0 to 20 cmH2O
- Mean Airway Pressure 0 to 120 cmH2O
- Inspiratory Time 0.2 to 4s
- Inspiratory:Expiratory Ratio 0.01 to 99.0
- Minute Ventilation-Controlled 0 to
99 L/min
- Minute Ventilation-Spontaneous 0
to 99 L/min
- Airway Temperature 15 to 40 deg. C
- Lung Compliance 10 to 150 mL/cmH2O
- Airway Resistance 1 to 60 cmH2O/L/s
- Leak 0 to 20 L/min
- Airway Flow Waveform -120 to +120
L/min
- Airway Pressure Waveform -20 to
+60 cmH2O
- Flow-Volume Graph
- Pressure-Volume Graph
- Minute Volume Wheel
Modes
- The Venturi® Ventilator's
flexible controls provide for a full range of
mandatory and spontaneous breaths.
- Assist Control, Pressure
Control, Variable Pressure Control
- Pressure Support/SIMV, CPAP/Assist
Control,
- Pressure Support/Variable Pressure
Control,
- Variable Pressure Support/Variable
Pressure Control,
- CPAP/SIMV, Variable Pressure
Support/SIMV,
- Apnea Ventilation
Electrical Requirements
- Voltage 115 VAC, 6 A or 230 VAC, 3
A
- Leakage Current <300 mA @ 115 VAC
- Power Cord 125 VAC, hospital grade, UL and CSA approved,
10 ft length.
- Internal Batteries Lead Acid 24 V,
18 A-h
Dimensions
- Height 52 in. (132 cm)
- Width 16 in. (41 cm)
- Depth 24 in. (61 cm)
- Weight 170 lbs. (77 kg)
|